Mindray DS USA, Inc. Has Initiated A Voluntary Recall Affecting The A3/A5 Anesthesia Delivery System

MAHWAH, N.J., Nov. 14, 2012 /PRNewswire-FirstCall/ -- Mindray Medical International Limited (NYSE: MR), a leading developer, manufacturer and marketer of medical devices worldwide, has initiated a voluntary recall affecting the A3/A5 Anesthesia Delivery System.  Mindray has initiated this recall due to the possibility of a system leak resulting from improper seating of the CO 2 absorbent canister gasket.

Should a system leak occur during use, fresh gas flow may be increased to compensate for any pressure loss resulting from the leak while the absorbent canister is unlocked and locked again to reseat the gasket. If the CO 2 absorbent canister gasket is improperly seated, however, a potential leak will likely present itself during the Automatic Circuit Leak and Compliance Test performed at startup and the Manual Leak Test recommended to be performed before each case, at which time the absorbent canister can be unlocked and locked again to reseat the gasket.

The canister gasket subject to this recall may be identified by presence of a small "step" in the gasket surface.  A3/A5 units with this gasket may continue to be used.  The proper seating of the canister gasket should be confirmed prior to use and after CO 2 absorbent is changed.

Customers who have a system(s) subject to this recall were notified by on August 8, 2012 by letter via US Postal Mail, return receipt required. To date, approximately 70% of units affected by this action have been corrected.  The issue is corrected through a replacement of the canister gasket by a Mindray Service or authorized representative.

Units with the affected canister gaskets were shipped between May 31, 2011 and July 15, 2012 in the United States, Latin America and Australia.

There have been no reports of injuries associated with this issue. Mindray became aware of the issue when a system leak was reported by a customer.  The cause of the leak was determined to be the result of an improperly seated canister gasket.

1 of 2

Check Out Our Best Services for Investors

Action Alerts PLUS

Portfolio Manager Jim Cramer and Director of Research Jack Mohr reveal their investment tactics while giving advanced notice before every trade.

Product Features:
  • $2.5+ million portfolio
  • Large-cap and dividend focus
  • Intraday trade alerts from Cramer
Quant Ratings

Access the tool that DOMINATES the Russell 2000 and the S&P 500.

Product Features:
  • Buy, hold, or sell recommendations for over 4,300 stocks
  • Unlimited research reports on your favorite stocks
  • A custom stock screener
Stocks Under $10

David Peltier uncovers low dollar stocks with serious upside potential that are flying under Wall Street's radar.

Product Features:
  • Model portfolio
  • Stocks trading below $10
  • Intraday trade alerts
14-Days Free
Only $9.95
14-Days Free
To begin commenting right away, you can log in below using your Disqus, Facebook, Twitter, OpenID or Yahoo login credentials. Alternatively, you can post a comment as a "guest" just by entering an email address. Your use of the commenting tool is subject to multiple terms of service/use and privacy policies - see here for more details.
Submit an article to us!
SYM TRADE IT LAST %CHG

Markets

DOW 17,826.30 -279.47 -1.54%
S&P 500 2,081.18 -23.81 -1.13%
NASDAQ 4,931.8150 -75.9760 -1.52%

Partners Compare Online Brokers

Free Reports

Top Rated Stocks Top Rated Funds Top Rated ETFs